An adaptive model approach for quantitative wrist rigidity evaluation during deep brain stimulation surgery

Thumbnail Image
Date
2016
Authors
Assis,S
Costa,P
Rosas,MJ
Vaz,R
João Paulo Cunha
Journal Title
Journal ISSN
Volume Title
Publisher
Abstract
Intraoperative evaluation of the efficacy of Deep Brain Stimulation includes evaluation of the effect on rigidity. A subjective semi-quantitative scale is used, dependent on the examiner perception and experience. A system was proposed previously, aiming to tackle this subjectivity, using quantitative data and providing real-time feedback of the computed rigidity reduction, hence supporting the physician decision. This system comprised of a gyroscope-based motion sensor in a textile band, placed in the patients hand, which communicated its measurements to a laptop. The latter computed a signal descriptor from the angular velocity of the hand during wrist flexion in DBS surgery. The first approach relied on using a general rigidity reduction model, regardless of the initial severity of the symptom. Thus, to enhance the performance of the previously presented system, we aimed to develop models for high and low baseline rigidity, according to the examiner assessment before any stimulation. This would allow a more patient-oriented approach. Additionally, usability was improved by having in situ processing in a smartphone, instead of a computer. Such system has shown to be reliable, presenting an accuracy of 82.0% and a mean error of 3.4%. Relatively to previous results, the performance was similar, further supporting the importance of considering the cogwheel rigidity to better infer about the reduction in rigidity. Overall, we present a simple, wearable, mobile system, suitable for intra-operatory conditions during DBS, supporting a physician in decision-making when setting stimulation parameters. © 2016 IEEE.
Description
Keywords
Citation