Please use this identifier to cite or link to this item: http://repositorio.inesctec.pt/handle/123456789/4384
Title: Strain sensor based on hollow microsphere Fabry-Perot cavity
Authors: Catarina Silva Monteiro
Susana Oliveira Silva
Orlando Frazão
Issue Date: 2017
Abstract: Fusion splicing technique was explored for the fabrication of two sensing structures based on hollow microsphere Fabry-Perot cavity. The first sensor proposed was fabricated with a hollow microsphere tip, working as a probe sensor. This structure was studied for lateral load pressure, yielding a 1.56 +/- 0.01 nm/N sensitivity. The second sensing structure relied on an in-line hollow microsphere, which allowed the detection of lateral load, with a sensitivity of 2.62 +/- 0.02 nm/N. Furthermore, the proposed structure enabled strain sensing, with a sensitivity of 4.66 +/- 0.03 pm/mu epsilon. The two sensing structures were subjected to temperature, presenting low thermal cross-sensitivity.
URI: http://repositorio.inesctec.pt/handle/123456789/4384
http://dx.doi.org/10.1117/12.2275862
metadata.dc.type: conferenceObject
Publication
Appears in Collections:CAP - Articles in International Conferences
CSIG - Articles in International Conferences

Files in This Item:
File Description SizeFormat 
P-00N-0Z5.pdf704.46 kBAdobe PDFView/Open