Aptamer-based fiber sensor for thrombin detection

Thumbnail Image
Date
2016
Authors
Luís Carlos Coelho
José Almeida
José Luís Santos
Pedro Jorge
Martins,MCL
Diana Viegas
Queiros,RB
Journal Title
Journal ISSN
Volume Title
Publisher
Abstract
The detection of thrombin based on aptamer binding is studied using two different optical fiber-based configurations: long period gratings coated with a thin layer of titanium dioxide and surface plasmon resonance devices in optical fibers coated with a multilayer of gold and titanium dioxide. These structures are functionalized and the performance to detect thrombin in the range 10 to 100 nM is compared in transmission mode. The sensitivity to the surrounding refractive index (RI) of the plasmonic device is higher than 3100 nmRIU(-1) in the RI range 1.335 to 1.355, a factor of 20 greater than the sensitivity of the coated grating. The detection of 10 nM of thrombin was accomplished with a wavelength shift of 3.5 nm and a resolution of 0.54 nM. (C) 2016 Society of Photo-Optical Instrumentation Engineers (SPIE)
Description
Keywords
Citation